Open original FDA record ↗Recall record
- FDA identifier
- Z-1235-2009
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 28, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect and/or incomplete data: Several potential safety issues while working with the syngo Image product including, but not limited to, data loss, errors in patient data due to different patient IDs, delays in image availability, inaccurate results during image measurements, incorrect review or diagnosis of displayed or printed images, measurement inaccuracies from the host modality and image
Source facts
- Action
- syngo Imaging has issued an Addendum to the User Manual, which was distributed to affected customers via Update Instructions IM118/08/S in November, 2008. Two safety related hotfixes (VB20G_HF4 and VB30A_HF26) will be released to those affected only after Update Instructions IM118/08/S has been performed. The addendum informs the user about safety relevant notes which pertain to syngo Imaging versions VB20G and VB30A. Update is a prerequisite for the following safety related hotfixes: VB20G_HF4 and VB30A_HF26.
- Initiated
- 2009-01-12
- Root cause
- Software design
- Terminated
- 2009-09-23
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1235-2009
- Recall status
- Terminated
- CDRH record ID
- 80254
- FDA classified
- 2009-04-28
- Recall event number
- 51423