Skip to content

Public FDA record

Z-1235-2014: Buzzaround XL scooters' Product Usage: mobility

Recall record Z-1235-2014 for Golden Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1235-2014
Company in FDA record
Golden Technologies, Inc.
FDA classification date
Mar 20, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Buzzaround XL scooters' front to rear lockup can become unintentionally disengaged due to possible improper fit of the front alignment cup to the rear alignment screw.

Source facts

Action
Golden Technologies sent a recall notification letter, dated January 14, 2014 to providers/dealers. A second letter, dated January 21, 2014 was sent to these same customers. Based on a slow initial response, the firm intends to send out the correction kit with the inspection acknowledgement form as soon as possible. Based on the overall response from dealers, a third consumer letter was drafted (notification3) for mailing directly to known consumers, if necessary. For help contact a special technical service hotline at 1-800-624-6374 ext. 392.
Initiated
2014-01-21
Root cause
Device Design
Terminated
2015-07-23
Distribution
Worldwide Distribution - USA Nationwide
Recall number
Z-1235-2014
Recall status
Terminated
CDRH record ID
125418
FDA classified
2014-03-20
Recall event number
67416

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗