Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures.
In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks.
Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.
Source facts
Action
Customer Notification Letter/Response Form is being sent to all affected consignees.
Initiated
2015-05-27
Root cause
Process control
Terminated
2017-02-23
Distribution
Terumo CVS has identified 33 domestic consignees and 0 international consignee that are affected by this recall
Recall number
Z-1235-2017
Recall status
Terminated
CDRH record ID
153034
FDA classified
2017-02-22
Recall event number
76386