Open original FDA record ↗Recall record
- FDA identifier
- Z-1235-2023
- Company in FDA record
- Sensus Healthcare, Inc.
- FDA classification date
- Mar 8, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.
Source facts
- Action
- Sensus healthcare issued an "Urgent Device Correction" to its consignees on 02/10/2023 by email. The notice requested the following actions: "Do not initiate beam delivery after a pause & resume sequence without verifying the amount of dose delivered. Do evaluate any patient treatments with version 1.9.2 that followed the pause & resume workflow. Do verify that the amount of dose to be delivered is as expected after each pause & resume sequence." The firm has developed a permanent software resolution for this issue. Customer service will contact the customer to coordinate upgrading to this revised software. Questions: Contact Sensus Customer Support at support@sensushealthcare.com or 561-578-4982
- Initiated
- 2023-02-10
- Root cause
- Software design
- Terminated
- 2024-01-31
- Distribution
- US Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.
- Recall number
- Z-1235-2023
- Recall status
- Terminated
- CDRH record ID
- 198892
- FDA classified
- 2023-03-08
- Recall event number
- 91795