Open original FDA record ↗Recall record
- FDA identifier
- Z-1236-2015
- Company in FDA record
- Zipline Medical
- FDA classification date
- Mar 2, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The manufacturers seal may not be present on the Nylon pouch resulting in lack of assurance for sterility.
Source facts
- Action
- Zipline sent an Urgent Voluntary product Recall letters to customers on February 5, 2015 by e-mail with possible follow up on site by firm representatives. The letter identified the affected product, problem and actions to be taken. Customers are asked to quarantine inventory immediately and examine the affected lots for complete pouch seals. Any product found with an incomplete seal should be destroyed or returned to ZipLine Medical. Please complete the enclosed response form as soon as possible, even if you no longer have stock of the lot numbers subject to this correction. This information is required to reconcile the inventory. Please contact Customer Service at 1-888-326-8999 to coordinate any product returns, replacement stock, or credit notes.
- Initiated
- 2015-02-05
- Root cause
- Packaging process control
- Terminated
- 2015-04-10
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Panama, Tehran, Ireland, Sweden, Japan, Hong Kong, Finland, Pakistan, Taiwan, Malaysia.
- Recall number
- Z-1236-2015
- Recall status
- Terminated
- CDRH record ID
- 133841
- FDA classified
- 2015-03-02
- Recall event number
- 70466