Skip to content

Public FDA record

Z-1236-2019: Atellica IM aTG (500 test kit) Siemens Material Number (SMN): 10995462

Recall record Z-1236-2019 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1236-2019
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA record creation date
Apr 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).

Source facts

Action
Siemens issued Urgent Medical Device Corrections, to Siemens Healthcare Diagnostics customers in the United States on December 7, 2018 delivered to customers on December 10, 2018. Urgent Field Safety Notices sent out to the country organizations on December 7, 2018 for distribution to customers outside the United States. These letters inform customers that Siemens Healthcare Diagnostics identified a bias with the affected lots of ADVIA Centaur aTG and Atellica IM aTG and action to take and resolution of the issue with new lots of the reagents. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Initiated
2018-12-07
Root cause
Nonconforming Material/Component
Terminated
2020-04-27
Distribution
Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana
Recall number
Z-1236-2019
Recall status
Terminated
CDRH record ID
171752
FDA record created
2019-04-27
Recall event number
82521

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗