Skip to content

Public FDA record

Z-1236-2023: CD5 Series of Genesis Reusable Rigid Sterilization Container Systems catalog number and product name: 1.

Recall record Z-1236-2023 for Carefusion 2200 Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1236-2023
Company in FDA record
Carefusion 2200 Inc
FDA classification date
Mar 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product did not meet shelf-life testing requirements resulting in a breach of the sterility of the Genesis container holding surgical equipment.

Source facts

Action
The recalling firm issued letters to customers and a separate letter to distributors dated 1/30/2023 via email and FedEx. The customer letters described the problem, potential clinical impact, provided product identification including photographs, and provided actions the consignees are to take. Instructions included: (1) Discontinue use of the affected product. Check all inventory locations within their institution for the affected product and immediately destroy all remaining product in their possession. A list of the components and their material makeup are provided so the consignee can have them disposed of according to their local facility's process; (2) Share the notification with all users of the affected product within their facility network to ensure they are also aware of the recall; (3) If the product was purchased from a distributor, the consignee is to contact the distributor directly for further instructions and credit resolution; (4) Complete the attached Customer Response Form and return it to the BD contact noted on the form. The letter also informs the consignee they will no longer manufacture or distribute the Genesis Reusable Rigid Sterilization Containers (CD5 Series only). The distributor letters were similar except they were requested to identify their customers who purchased the affected product and provide a copy of the customer letter to them. The distributors were to return the Distributor Response Form to the recalling firm acknowledging receipt and that they have read and understand the recall notice.
Initiated
2023-01-30
Root cause
Packaging change control
Distribution
Worldwide distribution - US Nationwide, including Puerto Rico. There was also government/military distribution. The countries of Australia, Canada, Denmark, Finland, Hong Kong, Ireland, Japan, Malaysia, Philippines, Poland, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom, United Arab Emirates.
Recall number
Z-1236-2023
Recall status
Open, Classified
CDRH record ID
198206
FDA classified
2023-03-09
Recall event number
91572

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗