Open original FDA record ↗Recall record
- FDA identifier
- Z-1237-2008
- Company in FDA record
- International Technidyne Corp.
- FDA classification date
- Aug 6, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect reading: ITC discovered that instruments built or repaired since approximately April 2007 were inadvertently left with the diagnostic mode enabled when shipped to customers.
Source facts
- Action
- ITC sent Urgent - Medical Device Field Recall - ITC No. 08-001 letters dated 2/08/2008 to hospitals and/or medical facilities identified as having affected instruments. The letter instructed users not to use the instrument if the diagnostic mode is enabled. ITC will reset affected devices and return them to the customers.
- Initiated
- 2008-02-13
- Root cause
- Process design
- Terminated
- 2009-07-24
- Distribution
- Worldwide Distribution: USA, United Arab Emirates, Malaysia, Saudi Arabia, India, Germany, United Kingdom, Greece, Romania, Singapore, ROC, Pakistan, Netherlands, Korea, France, Belgium, Switzerland, Canada, Thailand, Slovenia, Turkey and Australia.
- Recall number
- Z-1237-2008
- Recall status
- Terminated
- CDRH record ID
- 68273
- FDA classified
- 2008-08-06
- Recall event number
- 46773