- FDA identifier
- Z-1237-2010
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 5, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Due to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button. To remedy the issue described above, a revised cabling has been installed in all potentially affected syste
Source facts
- Action
- The recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.
- Initiated
- 2010-03-04
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-07-05
- Distribution
- The product was shipped to medical facilities nationwide.
- Recall number
- Z-1237-2010
- Recall status
- Terminated
- CDRH record ID
- 89653
- FDA classified
- 2010-04-05
- Recall event number
- 54999