Skip to content

Public FDA record

Z-1237-2010: Axiom Artis Systems operating with A100 Generator

Recall record Z-1237-2010 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1237-2010
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Apr 5, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Due to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button. To remedy the issue described above, a revised cabling has been installed in all potentially affected syste

Source facts

Action
The recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.
Initiated
2010-03-04
Root cause
Nonconforming Material/Component
Terminated
2012-07-05
Distribution
The product was shipped to medical facilities nationwide.
Recall number
Z-1237-2010
Recall status
Terminated
CDRH record ID
89653
FDA classified
2010-04-05
Recall event number
54999

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗