Open original FDA record ↗Recall record
- FDA identifier
- Z-1237-2026
- Company in FDA record
- Tornier, Inc
- FDA classification date
- Feb 2, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
Source facts
- Action
- On December 19, 2025, the firm began notifying affected customers via Urgent Field Safety Notice letters. Customers were instructed to immediately check their internal inventory for affected product. Firm stated that return of product is not necessary because the issue is detectable pre-operatively. The physical product itself is correctly marked, and after unpacking, should only be assigned to the correct place in the screw caddy. The notification is being provided for awareness only.
- Initiated
- 2025-12-19
- Root cause
- Labeling mix-ups
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1237-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217664
- FDA classified
- 2026-02-02
- Recall event number
- 98123