Open original FDA record ↗Recall record
- FDA identifier
- Z-1238-2016
- Company in FDA record
- Cook Inc.
- FDA classification date
- Mar 24, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Cook Medical is voluntarily recalling specific lots of the Ultraxx" Nephrostomy Balloon Set and the Ascend AQ Ureteral Dilation Balloon Catheter Set because of a labeling mix-up. Cook Medical has received reports of packages labeled as Ultraxx" Nephrostomy Balloon Set actually containing the Ascend AQ Ureteral Dilation Balloon Catheter Set and vice versa.
Source facts
- Action
- Cook Medical sent an Urgent Medical Device Recall Notification letter dated January 27, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers were instructed to quarantine any unused affected product. Immediately collect and return all unused affected products to Cook Medical If you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235. For questions regarding this recall call 800-457-4500.
- Initiated
- 2016-01-27
- Root cause
- Labeling mix-ups
- Terminated
- 2017-05-17
- Distribution
- Worldwide Distribution - US including GA, ID, LA, OR, SC, TX and Internationally to CHINA.
- Recall number
- Z-1238-2016
- Recall status
- Terminated
- CDRH record ID
- 144044
- FDA classified
- 2016-03-24
- Recall event number
- 73398