Reason stated in the source

Certain partial knee implant spacer blocks do not fit in the alignment tower instrument or handle.

Source facts

Action
Zimmer Biomet sent an Urgent Medical Device Correction Recall letter dated November 29, 2017. RISK MANAGER PHASE I: 1. Review this notification and notify affected personnel of the contents. 2. Product does not need to be returned at this time. You will receive a follow-up notice when the removal phase is initiated (Phase 2). If you identify a spacer block that will not assemble with these instruments, and you are unwilling to use the spacer block without this functionality, the spacer block may be returned to your sales representative. 3. Complete Certificate of Acknowledgement (Phase 1) and send to CorporateQuality.PostMarket@zimmerbiomet.com. SURGEON PHASE I: 1. Review this notification for awareness of the contents. 2. The spacer block may be used without the alignment tower and inserted and removed by hand from the joint, without the Persona Partial Knee handle. If you identify a spacer block that will not assemble with these instruments, and you are unwilling to use the spacer block without this functionality, the spacer block may be returned to your sales representative. 3. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 4. Complete Certificate of Acknowledgement (Phase 1) and send to CorporateQuality.PostMarket@zimmerbiomet.com. DISTRIBUTOR PHASE I: 1. Review this notification and notify affected team of the contents. 2. Product does not need to be returned at this time. You will receive a follow-up notice when the removal phase is initiated. 3. Should you or your account prefer to perform a functional check, this may be done by using the instructions in the surgical technique for assembling the alignment tower and handle to the spacer block. If you or your account identify a spacer block that will not assemble with these instruments, and the user is unwilling to use the spacer block without this functionality, the spacer block may be returned as a complaint to Pr
Initiated
2017-11-29
Root cause
Nonconforming Material/Component
Terminated
2019-04-05
Distribution
USA (nationwide) Distribution to the states of : AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI., and Internationally to Japan and Netherlands.
Recall number
Z-1238-2018
Recall status
Terminated
CDRH record ID
162188
FDA record created
2018-03-28
Recall event number
79383

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗