Skip to content

Public FDA record

Z-1238-2022: PX200 Emergency Relief Bed; Model No. PXEXEB362280A6TTZZH4C19

Recall record Z-1238-2022 for Oakworks Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1238-2022
Company in FDA record
Oakworks Inc
FDA record creation date
Jun 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product label was printed without the UDI number and warning label was printed with the incorrect weight capacity. The correct weight limit is 400 lbs., but the label states 500 lbs.

Source facts

Action
On April 21, 2022, the firm notified customers via mail and email of the field action. Customers should identify and locate all affected products. Distributors should quarantine the product and notify any customers who have received the product. For users, the product remains functional and can continue to be used. All customers should complete the included response form and return to Oakworks. Oakworks will set up a service visit to replace the labels on the affected beds. If you have any questions or concerns, please contact Oakworks Customer Service Department at 800-558-8850 or 717-235-6807 Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time.
Initiated
2022-04-21
Root cause
Error in labeling
Terminated
2024-09-19
Distribution
US Nationwide distribution in the states of MA, NJ, and PA.
Recall number
Z-1238-2022
Recall status
Terminated
CDRH record ID
193655
FDA record created
2022-06-06
Recall event number
90172

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗