Z-1238-2026: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G

Recall record Z-1238-2026 for Medtronic MiniMed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1238-2026
Company in FDA record
Medtronic MiniMed, Inc.
FDA classification date
Feb 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.

Source facts

Action
On 11/02/2025, the firm sent an initial "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that an issue was identified which may cause the MiniMed 780G pump to display an alarm indicating insulin delivery is being stopped (Pump error 43 or 41). Customers are instructed to: 1.Respond to all alerts/alarms from your pump. 2.Be prepared to complete the Reservoir & Set procedure after responding to a Pumperror 43 or Pump error 41 alarm. 3.Check that your Alert on High is set during the night or at times when you may not be actively monitoring the pump, making it easier to detect when insulin delivery remains suspended. 4. Keep the pump s audio enabled, making it easier to detect when insulin delivery remains suspended after an alarm is acknowledged. For questions or further assistance, contact Medtronic 24-Hour Technical Support line at 1-800-646-4633, option 1.
Initiated
2025-11-02
Root cause
Process design
Distribution
Worldwide - U.S. National distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The country of Israel.
Recall number
Z-1238-2026
Recall status
Open, Classified
CDRH record ID
217402
FDA classified
2026-02-02
Recall event number
97997

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗