Z-1239-06: Aksys PHD Personal Hemodialysis System for daily home hemodialysis; an automated high permeability hemodi

Recall record Z-1239-06 for Aksys, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1239-06
Company in FDA record
Aksys, Ltd.
FDA classification date
Jul 13, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The hemodialysis device was being marketed with a modified treatment length that exceeds the treatment length for which the device was originally designed.

Source facts

Action
Aksys sent a Clinical Bulletin: Field Correction Recall, CB Number: 048, dated 6/26/2006, via overnight Federal Express to all clinics or patients currently using the PHD instrument. CB #048 advises patients that until field modifications to the PHD System can be performed to limit the maximum treatment time to 170 minutes, Aksys recommends that users not exceed this treatment time in the interim. Each user was requested to contact his/her clinic to discuss how this change may affect treatment options and to complete and mail an enclosed pre-paid, self-addressed receipt acknowledgement card. A revised PHD System Operatror''s Manual was issued to all PHD System users on 7/12/06, which reflects the reduction in the maximum treatment time.
Initiated
2006-06-26
Root cause
Other
Terminated
2008-05-09
Distribution
Nationwide (California, Connecticut, Florida, Georgia, Illinois, Maryland, Massachusetts, Michigan, Ohio, Nebraska, New Jersey, New York, Tennessee, Texas, Virginia, Washington and Wisconsin), and internationally to the United Kingdom.
Recall number
Z-1239-06
Recall status
Terminated
CDRH record ID
46686
FDA classified
2006-07-13
Recall event number
35723

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗