Skip to content

Public FDA record

Z-1240-2014: 'TORAY' Hydrocoat Foley Catheter 12 x10 4.00 mm Fr x mL Product Usage: Latex Lubricous Coated Foley Cath

Recall record Z-1240-2014 for Convatec Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1240-2014
Company in FDA record
Convatec Inc.
FDA classification date
Mar 20, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The lumen of the Medline Sterile Foley Catheter was found to be blocked by excess lubricious coating material.

Source facts

Action
Convatec sent Urgent: Voluntary Medical Device Recall letters dated 2/7/2014 via UPS and email. Foreign customers were provided the notice via DHL. The letter identified the affected product, problem and actions to be taken. Customers are asked to perform a count of recalled product in inventory. All consignees should be contacted and be instructed to return product based on their procedures. Customers were instructed to complete and return the enclosed respoonse form.
Initiated
2014-02-07
Root cause
Device Design
Terminated
2014-12-18
Distribution
Worldwide Distribution and in the state of TX
Recall number
Z-1240-2014
Recall status
Terminated
CDRH record ID
126151
FDA classified
2014-03-20
Recall event number
67588

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗