The transformers installed on carts used for Richard Wolf medical video systems may overheat causing smoke and black plastic material to leak from the cart chassis.
Source facts
Action
Field Action Notice letters dated December 10, 2014 were sent to both direct accounts (customers). The letters instructed customers to: 1) Instruct the biomed to bypass the integrated power grid on the cart and not to use the electrical system on the affected carts until a qualified Omni Corporation service technician is able to inspect and repair the carts; and 2) Omni Corporation will schedule a qualified Omni field service technician, in conjunction with Richard Wolf and the customer's biomed unit, to visit and replace the transformer on the cart.
Initiated
2014-12-10
Root cause
Device Design
Terminated
2017-07-31
Distribution
Nationwide
Recall number
Z-1240-2015
Recall status
Terminated
CDRH record ID
133375
FDA classified
2015-03-03
Recall event number
70220