Skip to content

Public FDA record

Z-1240-2015: The I-Cart is used to store, transport, and power video equipment used in medical procedures.

Recall record Z-1240-2015 for Richard Wolf Medical Instruments Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1240-2015
Company in FDA record
Richard Wolf Medical Instruments Corp.
FDA classification date
Mar 3, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The transformers installed on carts used for Richard Wolf medical video systems may overheat causing smoke and black plastic material to leak from the cart chassis.

Source facts

Action
Field Action Notice letters dated December 10, 2014 were sent to both direct accounts (customers). The letters instructed customers to: 1) Instruct the biomed to bypass the integrated power grid on the cart and not to use the electrical system on the affected carts until a qualified Omni Corporation service technician is able to inspect and repair the carts; and 2) Omni Corporation will schedule a qualified Omni field service technician, in conjunction with Richard Wolf and the customer's biomed unit, to visit and replace the transformer on the cart.
Initiated
2014-12-10
Root cause
Device Design
Terminated
2017-07-31
Distribution
Nationwide
Recall number
Z-1240-2015
Recall status
Terminated
CDRH record ID
133375
FDA classified
2015-03-03
Recall event number
70220

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗