Open original FDA record ↗Recall record
- FDA identifier
- Z-1240-2018
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Mar 28, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer Biomet is conducting a medical device recall for the Persona Partial Knee Spacer Blocks in sizes 8mm, 9mm, 10mm, 12mm, and 14mm. Complaints received indicated that flexion has the potential to feel tight during range-of-motion trialing and that the final trial has the potential to appear looser in extension.
Source facts
- Action
- The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 10/31/2017 to customers titled "Affected Product: Persona Partial Knee Spacer Blocks Size 8mm, 9mm, 10mm, 12mm, 14mm". The letter described the product, problem, and actions to be taken. The letter instructed customers to review the notification and ensure that affected personnel are aware of the contents. Review inventory for affected product which would then be quarantined and ultimately removed by a firm sales representative. The customers were also instructed to complete and return the Certificate of Acknowledgement Form (even if you do not have affected products at your facility) via email to: CorporateQuality.PostMarket@zimmerbiomet.com or Fax to: 574-372-4265. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
- Initiated
- 2017-10-24
- Root cause
- Component design/selection
- Terminated
- 2019-04-03
- Distribution
- US Nationwide Distribution.
- Recall number
- Z-1240-2018
- Recall status
- Terminated
- CDRH record ID
- 162204
- FDA record created
- 2018-03-28
- Recall event number
- 79395