Z-1240-2019: Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips

Recall record Z-1240-2019 for Philips Medical Systems Nederlands. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1240-2019
Company in FDA record
Philips Medical Systems Nederlands
FDA record creation date
Apr 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling.

Source facts

Action
Philips issued an "URGENT - Medical Device Correction" letter on 3/25/2019 via certified mail to its customers. Philips Key Markets are responsible for distributing the letters outside of the U.S. Letter identifies the problem, health risk and actions to be taken. The customers were informed that no action from the customer or user of the device is required, as the Magnet Energization Device (MED) is located in a cabinet in the Technical Room, with the output terminals of the MED located behind locked protective covers. Only qualified servicing personnel have access to this. the Magnet PHILIPS Energization Device (MED) will be replaced by a new version, which does not have this problem. If you need any further information or support concerning this issue, please INFORMATION AND contact your local Philips representative or call the Quality Specialist DI BG Q&R Post Market Surveilla at 678-2960248 or email to: Etseguenet.Kinde@philips.com.
Initiated
2019-03-25
Root cause
Device Design
Terminated
2020-06-04
Distribution
US Distribution to state of: FL and Internationally to: Switzerland and Germany.
Recall number
Z-1240-2019
Recall status
Terminated
CDRH record ID
172033
FDA record created
2019-04-27
Recall event number
82610

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗