Recall record Z-1241-2025 for GE Healthcare (China) Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1241-2025
Company in FDA record
GE Healthcare (China) Co., Ltd.
FDA classification date
Feb 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Source facts

Action
On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2025-01-31
Root cause
Software design
Distribution
Domestic US distribution nationwide. International distribution worldwide.
Recall number
Z-1241-2025
Recall status
Open, Classified
CDRH record ID
212570
FDA classified
2025-02-28
Recall event number
96311

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗