Skip to content

Public FDA record

Z-1243-2007: Cerner Millennium RadNet Auto Launch Study and Auto Launch Report software functionalities.

Recall record Z-1243-2007 for Cerner Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1243-2007
Company in FDA record
Cerner Corp
FDA classification date
Sep 25, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Defects in the Auto Launch functionality make it possible for a mismatch of patient data.

Source facts

Action
The recalling firm initiated recall via telephone on 8/23/07 to their customers who had the 2007 version in their production environment, explained the reason for recall, and informed them they should turn off the affected functionality. The firm posted details of the situation on their website in two different documents (referred to as a Priority Review Flash), one dated 8/29/07 and the other dated 8/31/07. The 8/31/07 document was also mailed to their customers via certified mail on 9/4/07. The corrected software (EP28822 ver.1) was made available to the firm's clients on 9/7/07 via their website. Beginning 9/18/07, workarounds were discussed via telephone with the clients who had downloaded the software correction which contained a new bug that affected the dictation file. A new software correction to fix the dication file problem was posted on the recalling firm's website on 9/28/07 for clients to download and validate. A Priority review Flash #PR07-0176-2 client notification dated 10/4/07 was also posted on their website and mailed via certified mail return receipt.
Initiated
2007-08-23
Root cause
Software design
Terminated
2007-11-29
Distribution
Distribution was made to health care facilities located in LA, MO, NJ, MS, MI, NE, ND, and AR.
Recall number
Z-1243-2007
Recall status
Terminated
CDRH record ID
64345
FDA classified
2007-09-25
Recall event number
44739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗