Open original FDA record ↗Recall record
- FDA identifier
- Z-1244-2007
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Sep 26, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under certain circumstances, the product's software may crash and require a manual override or restart before functioning again. In addition, there is an unrelated addendum to the User Manual to clarify an icon not described in the previous User Manual.
Source facts
- Action
- As of 9/7/2007, Intuitive Field Service Engineers have performed the upgrade via service visit on 292 of 293 U.S. systems listed on the affected sites list. A notification letter that the upgrade has been performed was left with each site after the upgrade has been completed. The firm's distributors received advance notice of the upgrades 9/13/2007.
- Initiated
- 2007-05-03
- Root cause
- Software design
- Terminated
- 2008-07-02
- Distribution
- Worldwide, including USA, Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Romania, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela.
- Recall number
- Z-1244-2007
- Recall status
- Terminated
- CDRH record ID
- 52892
- FDA classified
- 2007-09-26
- Recall event number
- 38069