Open original FDA record ↗Recall record
- FDA identifier
- Z-1244-2008
- Company in FDA record
- AGFA Corp.
- FDA classification date
- Sep 11, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Java version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.
Source facts
- Action
- Agfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa software. A Mandatory Service Bulletin will inform and assist affected account in removing JAVA 1.6 if it has been applied and turning off the automatic updates via the JAVA Plug-in Control Panel.
- Initiated
- 2007-05-14
- Root cause
- Software change control
- Terminated
- 2010-12-09
- Distribution
- U.S. Nationwide and Canada.
- Recall number
- Z-1244-2008
- Recall status
- Terminated
- CDRH record ID
- 67348
- FDA classified
- 2008-09-11
- Recall event number
- 46384