Open original FDA record ↗Recall record
- FDA identifier
- Z-1244-2009
- Company in FDA record
- Migada Inc.
- FDA classification date
- May 6, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Syringe adaptor is disintegrating when attached to the end of the syringe.
Source facts
- Action
- A Field Safety Notice, dated 12/8/2008, was sent to all TEVADAPTOR customers advising them not to use drugs containing N,N-Dimethylacetamide. Teva Medical is presently revising the instruction leaflet supplied with each TEVADAPTOR to include a warning against the use of drugs containing N,N-Dimethylacetamide. Teva Medical will introduce this revised leaflet to current production as soon as practicable. In the meantime, Teva Medical will advise all users of the potential problem by means of the Urgent Field Safety Notice. Customers are to ensure that the safety notice is passed to all those who need to be aware within their organizations. Questions should be directed to their local contacts at B Braun, USA.
- Initiated
- 2008-12-18
- Root cause
- Component design/selection
- Terminated
- 2009-07-27
- Distribution
- Nationwide Distribution
- Recall number
- Z-1244-2009
- Recall status
- Terminated
- CDRH record ID
- 80247
- FDA classified
- 2009-05-06
- Recall event number
- 51421