Skip to content

Public FDA record

Z-1244-2009: TEVADAPTOR Closed Drug Reconstitution and Transfer System, a closed system used to prepare drugs, includi

Recall record Z-1244-2009 for Migada Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1244-2009
Company in FDA record
Migada Inc.
FDA classification date
May 6, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Syringe adaptor is disintegrating when attached to the end of the syringe.

Source facts

Action
A Field Safety Notice, dated 12/8/2008, was sent to all TEVADAPTOR customers advising them not to use drugs containing N,N-Dimethylacetamide. Teva Medical is presently revising the instruction leaflet supplied with each TEVADAPTOR to include a warning against the use of drugs containing N,N-Dimethylacetamide. Teva Medical will introduce this revised leaflet to current production as soon as practicable. In the meantime, Teva Medical will advise all users of the potential problem by means of the Urgent Field Safety Notice. Customers are to ensure that the safety notice is passed to all those who need to be aware within their organizations. Questions should be directed to their local contacts at B Braun, USA.
Initiated
2008-12-18
Root cause
Component design/selection
Terminated
2009-07-27
Distribution
Nationwide Distribution
Recall number
Z-1244-2009
Recall status
Terminated
CDRH record ID
80247
FDA classified
2009-05-06
Recall event number
51421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗