- FDA identifier
- Z-1245-2009
- Company in FDA record
- Sybaritic, Inc
- FDA classification date
- Jul 15, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sybaritic conducted a recall on the Hydration Station 806 because of the potential for hot water to spurt through steam ports into unit and cause burns. Also, potential for customers to inadvertently place their legs over the side wall of the unit directly onto the steam ports and cause burns. There were 18 injuries and 8 malfunctions reported.
Source facts
- Action
- Sybaritic, Inc. issued a letter dated May 15, 2007 addressed to "Hydration Station Customer". The letter informed the customer of the new 2007 Hydration Station Upgrade Improvement Kits including instructions and advised the customer to coordinate the installation of the components with their Technical Service Department. Further questions can be addressed to Sybaritic, Inc. Technical Administration at 1-800-445-8418 extension 373.
- Initiated
- 2007-05-15
- Root cause
- Device Design
- Terminated
- 2012-07-05
- Distribution
- Worldwide Distribution - US including states of AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MI, NE, NV, NH, NJ, NM, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, WV and WI and CANADA.
- Recall number
- Z-1245-2009
- Recall status
- Terminated
- CDRH record ID
- 81683
- FDA classified
- 2009-07-15
- Recall event number
- 51934