Z-1245-2009: Hydration Station with Pedestal (Part # E30003-01), Hydration Station with Pedestal & DVD (Part # E30003-

Recall record Z-1245-2009 for Sybaritic, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1245-2009
Company in FDA record
Sybaritic, Inc
FDA classification date
Jul 15, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Sybaritic conducted a recall on the Hydration Station 806 because of the potential for hot water to spurt through steam ports into unit and cause burns. Also, potential for customers to inadvertently place their legs over the side wall of the unit directly onto the steam ports and cause burns. There were 18 injuries and 8 malfunctions reported.

Source facts

Action
Sybaritic, Inc. issued a letter dated May 15, 2007 addressed to "Hydration Station Customer". The letter informed the customer of the new 2007 Hydration Station Upgrade Improvement Kits including instructions and advised the customer to coordinate the installation of the components with their Technical Service Department. Further questions can be addressed to Sybaritic, Inc. Technical Administration at 1-800-445-8418 extension 373.
Initiated
2007-05-15
Root cause
Device Design
Terminated
2012-07-05
Distribution
Worldwide Distribution - US including states of AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MI, NE, NV, NH, NJ, NM, NY, NC, OH, OR, PA, SC, TN, TX, UT, VA, WA, WV and WI and CANADA.
Recall number
Z-1245-2009
Recall status
Terminated
CDRH record ID
81683
FDA classified
2009-07-15
Recall event number
51934

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗