Skip to content

Public FDA record

Z-1245-2012: ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 35 cm Length Curved Jaw, Model # NSLG2C35 and ENSEAL L

Recall record Z-1245-2012 for Ethicon Endo-Surgery Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1245-2012
Company in FDA record
Ethicon Endo-Surgery Inc
FDA classification date
Mar 21, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Ethicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation: The ENSEAL device/system may continue to activate when the activation button is released prior to reaching the end of cycle (tone 3). (ii) No Activation

Source facts

Action
Ethicon Endo-Surgery sent an Urgent Medical Device Recall letter dated February 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately stop use of the affected product, fill out the attached Business Reply Form (BRF) and return to and return the affected product to: Ethicon Endo-Surgery ATT: ENSEAL G2 Recall 4545 Greek Road Cincinnati, Ohio 45242 Customers were instructed to choose one of the following response options: Return the Business Reply Form to their sales representative Call 1-800-873-3636, Option 6 Fax the BRF to 1-513-337-4138 For any questions regarding this recall call 513-337-3419.
Initiated
2012-01-26
Root cause
Nonconforming Material/Component
Terminated
2013-02-15
Distribution
Worldwide Distribution - USA including AZ, CA, CT, FL, IL, IN, KS, KY, MA, MI, MO, NC, NY, OH, PA, TN & TX and the countries of Arab Emirates, Austria, Belgium, France, Germany, Great Britain, Italy, Jordan, Kingdom of Saudi Arabia, Lebanon, Portugal, Slovenia, Sweden and Turkey.
Recall number
Z-1245-2012
Recall status
Terminated
CDRH record ID
107792
FDA classified
2012-03-21
Recall event number
61314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗