Open original FDA record ↗Recall record
- FDA identifier
- Z-1245-2013
- Company in FDA record
- Abbott Diabetes Care, Inc.
- FDA classification date
- May 9, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.
Source facts
- Action
- Abbott Diabetes Care sent an Urgent Product Recall letter dated April 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take one of the following options immediately to address the issue. 1) Customers can access a software update to resolve the issue at www.freestyleinsulinx.com/swupdate. 2) Customers can contact Abbott Diabetes Care Customers Service at 1-866-723-2697 to expedite return and replacement of the affected product at no charge. For questions customers were instructed to call Abbott Diabetes Care Customers Service at 1-866-723-2697. For questions regarding this recall call 510-239-2775.
- Initiated
- 2013-04-15
- Root cause
- Software design
- Terminated
- 2015-08-03
- Distribution
- Nationwide Distribution
- Recall number
- Z-1245-2013
- Recall status
- Terminated
- CDRH record ID
- 117228
- FDA classified
- 2013-05-09
- Recall event number
- 64876