Skip to content

Public FDA record

Z-1245-2017: Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application wh

Recall record Z-1245-2017 for Tosoh Smd Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1245-2017
Company in FDA record
Tosoh Smd Inc
FDA record creation date
Feb 22, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Power outages causes reporting software to shutdown.

Source facts

Action
Tosoh Bioscience sent a Customer Recall Notification letter dated December 21, 2016 to customers. The letter identified the affected product, problem and action to be taken. For questions call Tosoh Technical Support at 800-248-6764.
Initiated
2016-12-22
Root cause
Software design
Terminated
2019-06-18
Distribution
US Nationwide Distribution in the states of CT, MA, ME, MN, MO, NC, NE, NH, PA, SD, TX, VA, & WV.
Recall number
Z-1245-2017
Recall status
Terminated
CDRH record ID
152446
FDA record created
2017-02-22
Recall event number
76173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗