Skip to content

Public FDA record

Z-1245-2023: Fusion Bioline Vascular Graft, Part number M00201503046B0

Recall record Z-1245-2023 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1245-2023
Company in FDA record
Maquet Cardiovascular, LLC
FDA classification date
Mar 10, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One (1) Fusion Bioline Vascular graft from lot 25162546, which failed guideline straightness testing, was intermingled with lot 25162549 and distributed to customers. Use of the affected graft may result in reduction of blood flow or unwanted tension causing bleeding and the formation of a pseudo-aneurysm.

Source facts

Action
On January 30, 2023, the firm notified customers via Urgent Medical Device Removal letters. Customers are asked to examine your inventory immediately for lot 25162546. Please note, the Lot number appears on the device shelf boxes only. If you have the affected lot, please stop using and remove the device from areas of use. You are entitled to a replacement at no cost to your facility. You will receive replacement upon your acknowledgement that you have product for return. Distributors who have shipped this product should notify their customers. Please contact Getinge Customer Service at 888-943-8872 between the hours of 8:00 a.m. and 6:00 p.m. Eastern Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product.
Initiated
2023-01-30
Root cause
Process control
Distribution
US Nationwide distribution in the states of FL, IN, LA, MI, MO, NC, NV, NY, PA, TN, TX, VA, WA.
Recall number
Z-1245-2023
Recall status
Open, Classified
CDRH record ID
198468
FDA classified
2023-03-10
Recall event number
91671

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗