Skip to content

Public FDA record

Z-1246-2023: TFN-Advanced Femoral Nail 12MM/125 DEG TI CANN TFNA 200MM STERILE. Intended for temporary fixation and

Recall record Z-1246-2023 for Synthes (USA) Products LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1246-2023
Company in FDA record
Synthes (USA) Products LLC
FDA classification date
Mar 10, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

TFNA Femoral Nails labeled and etched as 200mm may measure 170mm in length, and TFNA Femoral Nails labeled and etched as 170mm may measure 200mm in length

Source facts

Action
DePuy Synthes issued Urgent Medical Device Recall (Removal) on 2/1/23 to US accounts. Canada and Japan and the affiliates in-country executing the recall per their local regulations. Letter states reason for recall, health risk, and action to take: Examine your inventory immediately to determine if you have the subject products and quarantinethem immediately. 2.Contact your DePuy Synthes Sales Consultant or customer support services at ra-dpyus-csspecial@its.jnj.com to coordinate the return/credit of the subject products. 3.Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) toOneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. Pleaseinclude in the email subject: FA 2219469 TFNA Incorrect Sized Nail 4.Please complete the attached Business Response Form even if you do not have the subject products onhand. 5.Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage,transport, store, stock, or use the subject products). 6.If any of the subject products have been forwarded to another facility, contact that facility and providethem with this notice. 7.Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information requests, please visit our website: https://www.jnjmedicaldevices.com/mir.
Initiated
2023-02-01
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of MS, OR, TX and the countries of Canada, Japan.
Recall number
Z-1246-2023
Recall status
Open, Classified
CDRH record ID
198609
FDA classified
2023-03-10
Recall event number
91702

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗