Open original FDA record ↗Recall record
- FDA identifier
- Z-1247-2009
- Company in FDA record
- Del Medical Systems Group
- FDA classification date
- Jun 30, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator.
Source facts
- Action
- Del Medical sent a "Field Correction Recall" letter with an acknowledgement form to the direct account dealers/assemblers on January 12, 2009 - February 4, 2009, informing them that required maintenance inspection of a retaining ring inside the column assembly may not be performed as required. If the retaining ring is not regularly inspected and replaced when worn or loose, there is the potential for the supporting cable to become detached from the column causing the column to fall and possibly contact the patient or operator. The accounts were requested to locate the affected systems and immediately take them out of service until correction by installing a Cable Mount Retention Retrofit Kit to eliminate the potential for the cable to become detached. Affected accounts were instructed to complete the enclosed acknowledgement form and return it via fax (847-288-7911) or mail to Del Medical. If there are any questions, please contact Del Medical Technical Support at 800-800-6006 or 847-288-7000.
- Initiated
- 2009-01-12
- Root cause
- Component design/selection
- Terminated
- 2010-06-17
- Distribution
- Worldwide Distribution -- US, Canada, Scotland and Pakistan.
- Recall number
- Z-1247-2009
- Recall status
- Terminated
- CDRH record ID
- 77462
- FDA classified
- 2009-06-30
- Recall event number
- 50841