Open original FDA record ↗Recall record
- FDA identifier
- Z-1247-2022
- Company in FDA record
- Medacta Usa Inc
- FDA record creation date
- Jun 7, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A Size 2 trial baseplate was assembled with Size 1 instrument sets.
Source facts
- Action
- The recalling firm emailed their sales agents on 4/13/2022 to notify them the implants were associated with a recall and must be returned immediately to the recalling firm. It also said that replacements are currently on order and will be overnighted to them once they are available. A copy of the Field Safety Notice from the manufacturer was forwarded to the sales agents. The Notice informed them the three affected lots have been assembled following an incorrect packing list, resulting in a size 2 trial baseplate being inserted in a size 1 instrument set, and that all the other devices included in the packing list and instrument set are correct.
- Initiated
- 2022-04-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2023-10-23
- Distribution
- Distribution was made to AK, CA, CT, IN, MA, NJ, and PA. There was no foreign/military/government distribution.
- Recall number
- Z-1247-2022
- Recall status
- Terminated
- CDRH record ID
- 193297
- FDA record created
- 2022-06-07
- Recall event number
- 90074