The intraoral x-ray systems were installed improperly with incorrect screws which allow the unit to pull away from the wall when the arm is fully extended. This condition may harm the patient.
Source facts
Action
Gendex sent a Urgent Medical Device Field Correction letter dated December 17, 2010 to the consignee, informing them that the x-ray units were installed with improper sized lag bolts and requesting that they contact Gendex Customer Service at 1-800-323-8029 to complete the field correction.
Initiated
2010-12-17
Root cause
Component design/selection
Terminated
2011-02-15
Distribution
California
Recall number
Z-1248-2011
Recall status
Terminated
CDRH record ID
96349
FDA classified
2011-02-15
Recall event number
57485