Z-1248-2025: ASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.

Recall record Z-1248-2025 for TERRAGENE S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1248-2025
Company in FDA record
TERRAGENE S.A.
FDA classification date
Feb 27, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Error in the IFU lists an incorrect incubation time on one occasion of 20 minutes rather than the correct incubation time of 60 minutes.

Source facts

Action
The recalling firm issued an email on 1/15/2025 to the main office of the distributor with an attached Distributor Customer Communication letter dated 1/15/2025, an Acknowledgement and Receipt Form, an updated IFU (Rev.B R01), and a User Letter for the distributor to issue to their customers. The letter to the distributor explains that all product is affected that contains version Rev.A R0 of the IFU because a discrepancy was identified in that version. The Distributor is to review their inventory records for the affected product, identify customers that received or may have received the devices and report it to the recalling firm through the provided Acknowledgement and Receipt Form via email. The Distributor is informed the product can be used according to the enclosed updated IFU (Rev.B R01). The Distributor was requested to notify their customer by using the enclosed User Letter and include the updated IFU with the User Letter. The Distributor is also informed they can continue distribution of their inventory by sending the digital user letter and the updated digital IFU with each delivery. An updated printed IFU will also be made available to users upon request. The User Letter was similar to the Distributor letter except that it instructs the user to discard older version of the IFU (Rev.A R0 and use the product according to the IFU attached to the letter and available on the firm's website. an Acknowledgement and Receipt Form was also enclosed for completion and return.
Initiated
2025-01-15
Root cause
Labeling False and Misleading
Distribution
US Nationwide distribution in the states of CA and KY.
Recall number
Z-1248-2025
Recall status
Open, Classified
CDRH record ID
212252
FDA classified
2025-02-27
Recall event number
96176

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗