Skip to content

Public FDA record

Z-1249-2025: Access Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily

Recall record Z-1249-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1249-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Feb 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.

Source facts

Action
Firm began notifying customers February 6, 2025. Customers were instructed to discontinue using and discard all remaining reagent packs of the Access EPO reagent lot identified. Customers are advised to share this recall with their medical director and perform a retrospective review of patient test results generated with this lot and assess the need for retesting.
Initiated
2025-02-03
Root cause
Unknown/Undetermined by firm
Distribution
Domestic distribution to AL AR AZ CA CT FL ID IL IN KS MA MI NC NE NJ NY OH OR PA TN TX UT WA WV. International distribution to Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Korea, Republic of, Netherlands, Philippines, Russian Federation, Saudi Arabia, Slovakia, Spain, United Kingdom of Great Britain and Northern Ireland
Recall number
Z-1249-2025
Recall status
Open, Classified
CDRH record ID
212434
FDA classified
2025-02-28
Recall event number
96255

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗