Z-1250-2019: Fetal Spiral Electrode, Model No. 9898 031 37631

Recall record Z-1250-2019 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1250-2019
Company in FDA record
Philips North America, LLC
FDA record creation date
May 2, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During use of the Philips FSE, it is possible for the metal electrode tip to break off and remain in the scalp of the newborn following labor and delivery.

Source facts

Action
On January 25, 2019, the firm distributed Customer Information Medical Device Correction letters to affected customers. The letter informed customers that there was an increase in complaints regarding the Fetal Spiral Electrode tip breaking off, preventing easy removal from the scalp of the newborn infant. Customers were advised to do the following: * Visually examine the FSE upon removal from the newborn scalp to verify the tip is intact. If it is not intact, appropriate measures to locate and remove the missing part should be taken. * Upon receipt of this notice, clinicians should review the following information from the device's Instructions for Use: - Do not pull the Spiral Tip from the fetal skin. Do not pull the FSE wires apart. - Do not over-rotate Spiral Tip during attachment. - Inspect the Spiral Tip to ensure that it is still attached to the FSE Hub. If the tip has separated from the hub and remains embedded in the presenting part, remove it using aseptic technique. Customers were provided with a reply form and asked to complete the form and return it to Philips: 1-877-499-7223 (eFax) or email to recall.response@philips.com. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377, option 2.
Initiated
2019-01-25
Root cause
Under Investigation by firm
Terminated
2020-12-08
Distribution
Distributed nationwide within US. Worldwide international distribution.
Recall number
Z-1250-2019
Recall status
Terminated
CDRH record ID
171593
FDA record created
2019-05-02
Recall event number
82422

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗