Z-1250-2025: Brand Name: TBS Osteo Product Name: TBS iNsight Model/Catalog Number: TBS iNsight v3.x / pn: OS-TBSi Soft

Recall record Z-1250-2025 for Medimaps Group. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1250-2025
Company in FDA record
Medimaps Group
FDA classification date
Feb 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential variability in calculations from fast array scans compared to array scans when operating on Hologic Horizon machines.

Source facts

Action
On February 5, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken by the user: Follow patient management recommendations. This notice needs to be passed on all organisation where the potentially affected devices have been transferred: the DXA manufacturer (Hologic), direct customers, distributors and field partners. Awareness on this notice and resulting action will be maintained through customer communications for an appropriate period of time until the end of the FSCA to ensure effectiveness of the corrective action. Customers are requested to report problem-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback.* Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Action Being Taken by the Manufacturer: Software upgrade Medimaps will issue a maintenance release of the software by the end of February in the worst case and deploy it progressively in the field. The exact plan will be further communicated to customers. The associated FSCA for strongly recommended software update should be completed in 6 months starting from the release of the update version at the end of February 2025.
Initiated
2025-02-03
Root cause
Software Design Change
Distribution
Domestic: AL, AZ, CA, CO, DE, FL, IL, LA, MA, MD, MI, MN, MO, MS, NC, NM, NY, OH, PA, SC, TX, UT, WA. International: ALGERIA, ANDORRA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, SOUTH KOREA, LEBANON, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, NEPAL, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY
Recall number
Z-1250-2025
Recall status
Open, Classified
CDRH record ID
212390
FDA classified
2025-02-28
Recall event number
96233

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗