New protocols for MR sequences may result in higher RF energies deposited during MR scanning, generating heat in uninsulated fixation posts.
Source facts
Action
Elekta sent Important Notice Affected Product: Fixation Posts, A331, dated June 22, 2009.
The Important Notice will be added to the Customer User manual upon receipt.
Initiated
2009-10-02
Root cause
Labeling design
Terminated
2011-02-17
Distribution
Nationwide
Recall number
Z-1251-2011
Recall status
Terminated
CDRH record ID
96757
FDA classified
2011-02-15
Recall event number
57604
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.