Open original FDA record ↗Recall record
- FDA identifier
- Z-1251-2015
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- Mar 6, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Improper installation of Tube arm, and Wall stand may cause it to fall down, Improper installation of table may cause it to float.
Source facts
- Action
- Philips Healthcare issued notice to foreign Distributors on 1/23/15 to issue and advise customers to stop using the device until a Philips Engineer checks the device. After the engineer verifies that the tube arm, Wall stand and table were properly assembled, the device may be used again. Philips will subsequently install a new Tube Arm and Wall stand fixing structure that is designed to prevent this hazard if the tube arm or Wall stand is incorrectly reinstalled in the future. Philips will subsequently provide the method to adjust the T-nuts of table to make sure it was properly assembled. A Philips Service Engineer will contact affected customers to install the solution.
- Initiated
- 2015-01-23
- Root cause
- Process control
- Terminated
- 2016-06-07
- Distribution
- Worldwide Distribution. Argentina, Australia, Burkina-Faso, Chile, China, Colombia, Ecuador, France, Germany, Hungary, Indonesia, Iraq (NL), Kuwait, Libya, Macedonia, Malaysia, Mayotte, Poland, Saudi Arabia, Slovenia, Spain, Switzerland, Turkey, United Kingdom, and Uzbekistan.
- Recall number
- Z-1251-2015
- Recall status
- Terminated
- CDRH record ID
- 133671
- FDA classified
- 2015-03-06
- Recall event number
- 70432