Z-1251-2022: NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Model No. 300900

Recall record Z-1251-2022 for NeuMoDx Molecular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1251-2022
Company in FDA record
NeuMoDx Molecular Inc
FDA record creation date
Jun 8, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for reduced analytical sensitivity for samples in the 1x - 3x LOD range for Flu A, RSV, and SARS-CoV-2 targets. This may lead to false negative results, allowing ongoing transmission to others and delaying appropriate treatment.

Source facts

Action
On May 13, 2022, the firm distributed Urgent Medical Device Correction letters to impacted customers. The letter informed customers of the reduced analytical sensitivity for Flu A, SARS-CoV-2, and RSV. The firm also specified that only the RSV target showed a similar reduction in clinical sensitivity; Influenza A and SARS-CoV-2 did not show a reduction in clinical sensitivity. Influenza B is not affected by this issue. The letter reminded users that negative results do not preclude infection with Influenza virus, RSV, and/or SARS-CoV-2, and the device should not be used as the sole basis for patient management and treatment decisions. Customers should share the notice with all affected users. If you have any questions regarding this recall, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com.
Initiated
2022-05-13
Root cause
Process control
Distribution
Domestic distribution to FL, NY, MD, PA, VA, MI, TN, NM. Foreign distribution to AT, AU, BE, CH, DE, FI, FR, GB, GR, HK, ID, IT, LT, NL, SA, SE, and SI.
Recall number
Z-1251-2022
Recall status
Open, Classified
CDRH record ID
194016
FDA record created
2022-06-08
Recall event number
90275

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗