Z-1251-2025: human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single

Recall record Z-1251-2025 for HUMAN MED AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1251-2025
Company in FDA record
HUMAN MED AG
FDA classification date
Feb 28, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.

Source facts

Action
The recalling firm issued a Field Safety Notice to Distributors dated 10/18/2025 via email explaining the reason for the Field Safety Notice, the potential hazard, and informing the consignee to discontinue delivering the affected products to customers and users and to dispose of the product if still in their warehouse. A Field Safety Notice was enclosed to be issued to the consignee's users and customers who received the affected product from the Distributor. The User and Customer letter dated 10/17/2024 was similar to the Distributor letter except provided instructions to check the spray pattern as instructed in the user manuals prior to using the cannula. If the pattern was not in a flat, fan-shaped pattern, they are not to use the cannula and it should be sorted out and disposed of properly. A confirmation form was enclosed to confirm the cannula has been correctly disposed of and the form was to be sent to the distributor or to the email address provided.
Initiated
2024-10-18
Root cause
Process control
Distribution
US Nationwide distribution in the state of FL.
Recall number
Z-1251-2025
Recall status
Open, Classified
CDRH record ID
212112
FDA classified
2025-02-28
Recall event number
96137

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗