Open original FDA record ↗Recall record
- FDA identifier
- Z-1252-2010
- Company in FDA record
- Picis Inc.
- FDA classification date
- Apr 6, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Under specific timing of conditions and in configuration with 3rd party infusion pumps, an error within the clinical application causes the manual documentation of clinical orders to be changed to an automatic update status. At this point, the order documentation becomes frozen at the last dose and/or rate and the user is unable to modify, stop or delete it. This error could impact the fluid bala
Source facts
- Action
- Picis issued letter notification beginning on March 8, 2010 with software correction information. The letter communicates details of the problem, along with recommendations for mitigating the potential to experience this error condition. Use of the product should be discontinued until the update described in the letter is performed. Customers with questions should call the company at 781-557-3000.
- Initiated
- 2010-03-08
- Root cause
- Software design
- Terminated
- 2013-07-01
- Distribution
- NJ Canada, Australia, Austria,Denmark, Finland, Germany Norway,Portugal, Sweden, France,Netherlands, Spain.
- Recall number
- Z-1252-2010
- Recall status
- Terminated
- CDRH record ID
- 89514
- FDA classified
- 2010-04-06
- Recall event number
- 54934