Z-1253-06: Ultra-Filtration (UF) Removal Regulators(Rgltr Assy UF REM Molded Spare); a spare part used with Baxter''

Recall record Z-1253-06 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1253-06
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Jul 25, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.

Source facts

Action
Baxter sent Urgent Device Correction letters dated 06/21/06 to all System 1000, TINA, Altratouch, Arena and Aurora Hemodialysis Instrument customers who may have received the nonconforming assemblies via first class mail on the same date, to the attention of the Hemodialysis Administrator. The letters informed the accounts that the three lots of UF Removal Regulators were potentially assembled incorrectly. While these regulators are identical in appearance to properly assembled components, the mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy. Replacement regulators were included with the recall letter, and the accounts were requested to return the regulators from their previous shipment to Baxter using the enclosed mailing materials. The accounts were requested to complete the enclosed reply form indicating the number of units received and the number of units returned and acknowledging receipt of the letter and dissemination of the information to their staff and to other services of facilities, as applicable, and fax the sheet to 1-727-544-3025. Any technical questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 3, option 2, choice 1.
Initiated
2006-06-21
Root cause
Other
Terminated
2008-03-04
Distribution
Worldwide Distribution including the states of Arizona, California, Connecticut, Florida, Georgia, Maryland, Missouri, Ohio, Oklahoma, Tennessee, Texas and Puerto Rico, and international distribution to Belgium, Canada, Thailand, Chile, Austrailia, Peru, Argentina, Panama and Mexico.
Recall number
Z-1253-06
Recall status
Terminated
CDRH record ID
46645
FDA classified
2006-07-25
Recall event number
35706

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗