Z-1253-2021: Software versions syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a in the Computed tomography systems as follows

Recall record Z-1253-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1253-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Mar 15, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

SOMATOM systems-issues in with software syngo.CT VA30A_SP2 or syngo.CT VA30A_SP2a, may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Delay in diagnosis or patient rescan may occur. Sporadic software errors during interventional workflows may also result

Source facts

Action
Siemens Medical Solutions USA, Inc., provided Customer Information Letter on 3/3/21 to affected customers via CT063/20/S. Customers with valid email addresses emailed a copy of the CSAN via Adobe Mega Sign. The CSAN will be distributed to the remaining customers address via hand delivery by the Siemens Customer Service Engineer. The letter states reason for recall, health risk and action to take: Siemens Healthineers has developed software update syngo.CT VA30A_SP3 to ensure uninterrupted scanning workflows and to reduce the number of user notifications. This update will also provide improvements related to the general system, examination and reconstruction processes, as well as improvements for dual energy post-processing. Following the corrective action, the cause has been eliminated and recurrence of the identified issues are prevented. This software update will be provided to you free of charge. How will the corrective action be implemented? Software update CT084/20/S will be performed onsite for SOMATOM X.cite (Model #11330001). For all other systems, the update CT064/20/S will be performed remotely. The remote software updates process will require approximately 180 minutes for completion. If you have any unresolved questions or you require technical support, please contact your local application specialists or your local service/sales organization at 1-800-888-7436.
Initiated
2021-03-03
Root cause
Software design
Terminated
2023-12-28
Distribution
US Nationwide distribution.
Recall number
Z-1253-2021
Recall status
Terminated
CDRH record ID
186057
FDA record created
2021-03-15
Recall event number
87454

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗