Open original FDA record ↗Recall record
- FDA identifier
- Z-1253-2025
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Mar 6, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
There is the potential for missing motor mounting screws, which may have occurred during the servicing process.
Source facts
- Action
- Baxter issued an "URGENT MEDICAL DEVICE RECAL" notice to its consignees on 02/05/2025 via US Mail. The notice explained the issue, hazard involved, and requested the following actions be taken: "1. Immediately locate, isolate, and cease all use of Spectrum pumps with the affected serial numbers. The product code and serial number can be found on the bottom of the infusion pump. 2. Contact Baxter to arrange for the return of the affected pumps for inspection and reservicing as applicable. 3. If you received this communication directly from Baxter, acknowledge receipt on the enclosed reply form even if you have no remaining inventory. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to all departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of the affected product that you distributed to customers and check the associated box on the customer portal. For general questions regarding this communication, or quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 Monday through Friday, between 7:00 am and 7:00 pm Eastern Time."
- Initiated
- 2025-02-05
- Root cause
- Under Investigation by firm
- Distribution
- US: Nationwide
- Recall number
- Z-1253-2025
- Recall status
- Open, Classified
- CDRH record ID
- 212492
- FDA classified
- 2025-03-06
- Recall event number
- 96289