Reason stated in the source
During ExacTrac Robotics installation on the Varian Exact Couch, the couch height position indication is re-calibrated to read out the correct positions. This re-calibration could cause the eventual failure of the vertical lift mechanism of the couch, which might result in patient injury or death.
Source facts
- Action
- BrainLab issued a Product Notification Letter to the hospitals on 6/2/06 to advise the users not to lower the couch below 50 cm, and sent a second letter to the users on 6/7/06, informing them that BrainLab would remove the ExacTrac Robotics from the Exact Couch and re-calibrate the couch to restore the software limit switch. The Robotics will be re-installed once a permanent solution is available.
- Initiated
- 2006-06-02
- Root cause
- Other
- Terminated
- 2008-02-24
- Distribution
- Nationwide, including the states of California, Colorado, Florida, Illinois, Maryland, Michigan, Missouri, New Hampshire, New York, North Carolina, Ohio, Pennsylvania, Tennessee and Texas
- Recall number
- Z-1255-06
- Recall status
- Terminated
- CDRH record ID
- 46713
- FDA classified
- 2006-07-27
- Recall event number
- 35742