Open original FDA record ↗Recall record
- FDA identifier
- Z-1255-2021
- Company in FDA record
- R & D Systems, Inc.
- FDA record creation date
- Mar 15, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.
Source facts
- Action
- Users of these kits are requested to stop using these kits for measuring Human sTfR for diagnostic purposes in patient samples and to discard all unused kits or kit components. Customers were instructed to complete and return the MEDICAL DEVICE RECALL RETURN RESPONSE Acknowledgement and Receipt Form to R&D Systems by February 20, 2021, by emailing the form to techserv@bio-techne.com. For questions regarding this recall, or need to obtain a product credit, contact Bio-Technes Technical Service Department at 1-612-379-2956 or techserv@bio-techne.com.
- Initiated
- 2021-01-20
- Root cause
- Under Investigation by firm
- Terminated
- 2023-02-13
- Distribution
- USA: (FL, LA, CA, OH), CHINA, UNITED KINGDOM, JAPAN, SINGAPORE, Jakarta Selatan,
- Recall number
- Z-1255-2021
- Recall status
- Terminated
- CDRH record ID
- 185569
- FDA record created
- 2021-03-15
- Recall event number
- 87223