Open original FDA record ↗Recall record
- FDA identifier
- Z-1256-2022
- Company in FDA record
- Steris Corporation
- FDA record creation date
- Jun 9, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Source facts
- Action
- On April 22, 2022, the firm distributed notification letters to affected customers. All affected customers will be contacted by a STERIS Service Technician to arrange an onsite inspection of their unit(s) and perform the correction. Users may continue to use their affected devices prior to completion of the correction. Users are reminded to follow instructions outlined in the Operator Manual should the unit alarm "Drying Temperature Too High" by aborting the cycle, positioning building electrical disconnect switch to OFF, and contacting STERIS. If you have questions regarding this issue, please contact Sarah Brown, Senior Product Manager at 440-392-7571, STERIS Customer Service at 1-800-548-4873, or your local STERIS Representative.
- Initiated
- 2022-04-22
- Root cause
- Component design/selection
- Terminated
- 2024-11-04
- Distribution
- Worldwide distribution - US Nationwide.
- Recall number
- Z-1256-2022
- Recall status
- Terminated
- CDRH record ID
- 193845
- FDA record created
- 2022-06-09
- Recall event number
- 90222