Incorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10% lower than what the displayed value should be when the scanner executes the exam Plan.
Source facts
Action
Toshiba America Medical Systems (TAMS) issued notification to consignees 9/7/2010, via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification.
Initiated
2010-09-07
Root cause
Software design
Terminated
2012-04-20
Distribution
Nationwide distribution.
Recall number
Z-1257-2011
Recall status
Terminated
CDRH record ID
96386
FDA classified
2011-02-15
Recall event number
57499