It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Source facts
Action
DJO Global sent an Urgent Field Safety Notice letter dated January 29, 2018, to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call (512) 834-6255
Initiated
2018-01-29
Root cause
Device Design
Distribution
US Nationwide Distribution
Recall number
Z-1257-2018
Recall status
Completed
CDRH record ID
162025
FDA record created
2018-03-28
Recall event number
79241